Continuµm User Guide
55
8. If you have an optional motorized stage, home the stage to
the area of interest in your sample.
Choose Move Stage from the Atlµs menu and then choose Set
Home to Current.
9. Refine the standard experiment background and sample
point area settings for your sample position.
Use the tool palette and Atlµs menu items as follows:
Click the Full Range View button to zoom the navigation pane
all the way out to display the full range of stage travel allowed
by your stage controller. You should now be able to see the
sample (crosshairs labeled with a number) and the background
point (crosshairs with the letter B) defined in the standard
experiment.
If you cannot, use the Zoom buttons to zoom in until you can see
the both sample and background point indicators.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...