Limiting stage travel
Limit stage travel, available for motorized stages through System
Configuration in the Atlµs menu, allows you to further control
horizontal movement of the stage. This is useful to prevent stage
collision damage when performing transmission experiments on
thick samples (greater than 2250 µm (0.09 in)) and/or with large
condensers (such as the Reflachromat condensers) installed. A
checkbox allows you to toggle between restricted travel and full
travel with a single mouse click. This feature is not available for
standard manual or glide stages.
While performing reflection experiments on samples that are
2250 µm
(0.09 in) or less in thickness, there in no need to use stage
travel limits. Simply lower the condenser to a level below the bottom
of the stage (or remove it all together) and the full 27000 µm x
76200 µm (5 in x 3 in) stage travel is available.
Notice
Use care and avoid collisions while setting the travel limits. Be sure
that the physical center of the stage is centered under the objective
and that the universal slide holder is removed from the stage before
opening the Limit Stage Travel dialog. Damage due to stage
collision is not covered under warranty.
To limit stage travel:
1. Click the Limit Stage Travel button in the Microscope field
of the System Configuration dialog box.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...