If you were analyzing a sample that is compressed between
windows or in a compression cell, you would set the objective
compensation ring to match the thickness of the top window.
If your sample was on a substrate or in a compression cell, you
would set the condenser compensation ring to match the
thickness of the substrate or bottom window of the cell.
If you were analyzing the sample with a fixed-compensation
objective and condenser, you would use an objective and
condenser with a refractive index compensation value of 0.