13. Verify the experiment parameters.
Select Experiment Setup from the Collect menu and verify the
experiment parameters, as follows. Close Experiment Setup
when you are finished.
Where
Parameter
Setting
Collect tab
Number of scans
16
Resolution
8
Final format
%Transmittance
Correction
None
Background handling Background before
every sample
Collect 64 scans for
background
Bench tab
Sample Compartment Left µscope %T
OR
Right µscope %T
Detector
your single-element
detector (Usually
MCT-A)
Beamsplitter
your compatible
beamsplitter
(usually KBr)
Source
IR
Accessory
None
Window
None
Bench tab
Sample Compartment Left (or right,
depending on your
setup) µScope, %T
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...