Symptom
Possible Cause
Solution
Video pane of the OMNIC
Atlµs or µView window is
dark grey or black,
continued…
A fuse is blown.
Check fuses and replace any fuse that is
blown.
See
“Replacing a fuse” in
“Microscope Help topics” in the OMNIC
Help menu.
The illuminator bulb(s) is
burned out.
Replace the bulb(s).
See
“Replacing an
illuminator bulb” in the “Service and
Maintenance” Help book, available through
Microscope Help Topics in the OMNIC Help
menu.
Microscope power is turned off
or power cord is disconnected.
Make sure power cord is connected and
power switch is turned on (“I” position).
See
“Turning on the microscope power” in
“Preparing your Microscope” chapter.
The microscope, camera,
computer, and/or video capture
card are malfunctioning.
Contact technical support.
Video pane of the OMNIC
Atlµs or µView window is a
solid blue color.
View selector is set for viewing
through the eyepieces only.
Set view selector to allow the image to pass
to the camera.
See
“Preparing the optical
components” in the “Preparing the
microscope” chapter.
Field iris is closed.
Open field iris for current mode.
See
“Adjusting field of view and image contrast”
in the “More about the Hardware” chapter.
The microscope power or
illuminator power is OFF.
Verify that the microscope and illuminator
power is ON.
See
“Turning on the
microscope power” in “Preparing the
Microscope” chapter.
The power source or power
cord is damaged.
Verify that the power cords are connected to
a working AC source (wall outlet or power
strip) and firmly seated at both ends. Replace
any damaged cables.
continued…
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...