Thermo Scientific
36
Reticle focus
and diopter adjustments
Each eyepiece is marked with a scale. When you finish focusing the
reticle and adjusting the diopter to match focus, take note of the
scale settings. In the future you will be able to adjust the diopter
quickly by turning the eyepiece to this reading.
1. Focus the reticle in the right eyepiece.
Hold the inner ring with one hand and turn the outer ring until
the reticle cross hairs are clear. Once the cross hairs are clear, use
the stage focus knob to obtain a sharp image of the pinhole
through the right eyepiece.