Continuµm User Guide 155
Symptom
Possible Cause
Solution
Video pane of the OMNIC
Atlµs or µView window is
dark grey or black,
continued…
Field of view and/or contrast
controls are improperly
adjusted.
Adjust the field and aperture iris settings.
See
“Adjusting field of view and contrast” in the
“More about the hardware” chapter.
Microscope and/or illuminators
are turned off.
Verify that the microscope and illuminator
power is ON.
See
“Turning on the
microscope power” in “Preparing the
Microscope” chapter.
Verify that the power cords are connected to
a working AC source (wall outlet or power
strip) and firmly seated at both ends.
Illuminators are improperly
adjusted.
Adjust the illuminator intensity.
See
“Preparing the optics” in “Preparing the
Microscope” chapter.
No video signal is reaching the
computer.
Verify that the microscope and illuminator
power is ON.
See
“Turning on the
microscope power” in “Preparing the
Microscope” chapter.
Verify that the power cords are connected to
a working AC source (wall outlet or power
strip) and firmly seated at both ends. Replace
any damaged cables.
Verify that the video cable (composite or S
Video) is connected between the camera and
the microscope (analog) or the USB cable is
connected from the microscope to the
computer (digital)..
Replace any damaged cables.
continued…
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...