Continuµm User Guide
13
4. After you have filled the funnel once, allow the detector
dewar to cool for one to two minutes before adding more
liquid nitrogen.
Filling the detector too quickly can cause unnecessary boil off
and hazardous splatter.
5. Continue filling the funnel and allowing the nitrogen to drain
into detector dewar.
The dewar is full when liquid nitrogen no longer drains into it.
6. Allow the detector temperature to stabilize.
For best results, allow the detector to stabilize for at least 20
minutes before analyzing samples.
Installing
the purge baffle
The purge baffle protects the optical components in the microscope
from moisture in your breath. The baffle deflects your breath away
from the polarizer slot when you do not have a polarizer installed.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...