4. Click the X-Y Stage button in the Microscope field.
Notice
Always remove the nosepiece and lower the condenser completely
before you initialize the stage. Damage resulting from stage
collisions is not covered by warranty.
5. When the Stage Initialization information box appears,
choose OK to begin calibrating horizontal stage movement.
Choose Cancel if you do not wish to calibrate the XY-axes at this
time.
6. Wait while the stage initializes.
While the stage is being calibrated, no other system
configuration parameters can be edited and no other system
configuration operations can be conducted.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...