Thermo Scientific
106
Notice
Excessive pressure can damage the ATR crystal. Be sure to stop
raising the stage in the next step when the upper indicator
illuminates.
11. Slowly raise the stage and stop immediately when the upper
Contact Alert indicator on the front panel illuminates.
This red indicator shows that the contact pressure of the sample
area against the ATR crystal is optimal for infrared analysis.
12. If you are using an ATR objective with a ZnSe or diamond
crystal, look through the viewer and make sure the sample
area of interest is positioned as desired in the field of view.
Good contact between the crystal and sample is indicated by a
darkening or “wetting” of the sample image.
If the area of interest is not positioned as desired, lower the stage
until the sample is slightly below the objective, reposition and
focus the sample with the selector in survey position, and then
return to step 10.
13. Choose Experiment Setup from the Collect menu, and verify
that the detector signal intensity is adequate.
The detector signal is shown in the live display on the Bench tab
of the Experiment Setup dialog box.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...