Continuµm User Guide 167
Symptom
Possible Cause
Solution
When the Atlµs window is
open, OMNIC does not
respond but the Windows
Task Manager does not list
OMNIC as “Not
Responding”.
Serial communications
problem.
Verify that the microscope and the stage
controller are powered on.
Verify that the microscope and stage
controller are connected to an AC source (a
working wall outlet or power strip).
Verify that the serial cable that connects the
microscope to the computer is attached
firmly seated at both ends.
Verify that the serial cable that connects the
stage controller to the computer is attached
firmly seated at both ends.
Verify that the cable that connects the stage
controller to the stage is firmly seated at both
ends.
Replace any damaged cables.
Verify that the serial ports identified in
System Configuration match the hardware
connections for the microscope and stage
controller.
Shutdown the computer, power OFF the
illuminators, microscope, stage controller,
spectrometer, and Contact Alert (if
installed). Wait 30 seconds and then power
ON the system using the sequence described
in “Turning on the microscope power” in the
“Preparing the Microscope” chapter.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...