Continuµm User Guide 105
9. Move the crystal into the beam path.
If you are using an ATR objective with a ZnSe or diamond
crystal type, slide the selector into the ATR position.
If you are using the slide-on ATR attachment, move the crystal
slide to the middle position.
If you are using an ATR objective with a Ge or Si crystal, you
can move the selector to ATR.
10. While watching the lower Contact Alert indicator on the
front panel, carefully and slowly raise the stage using the
stage focus knob until the indicator illuminates.
This green indicator shows that the sample has made initial
contact with the ATR crystal.