6. Set the following experiment parameters:
Where
Parameter
Setting
Collect tab
Number of scans
200
Resolution
4
Final format
%Transmittance
Correction
None
Background handling Background before
every sample
Collect 32 scans for
background
Bench tab
Sample Compartment Left µscope %T
OR
Right µscope %T
Detector
your single element
detector
Beamsplitter
your compatible
beamsplitter
Source
IR
Accessory
None
Window
None
Max range limit
2600
1
Min range limit
2500
2
Notes:
1. For InGaAs detectors set the Max range limit to 12000 cm
-1
.
2. For InGaAs detectors, set the Min range limit to 3800 cm
-1
.