4. Connect the cable from the sensor plate to the connector on
the Contact Alert Control box and set the sensor plate aside.
To prevent damage when the stage initializes, be sure that the
cable that connects the sensor plate to the microscope is behind
and under the stage.
5. Power on the microscope, stage controller, illuminators,
computer, monitor, printer, and other computer peripherals,
Contact Alert controller, and then, finally, power on the
spectrometer.
Be sure that at least 20 seconds elapses between powering on the
microscope and powering on the spectrometer.
See
“Turning on power” in the “Preparing the Microscope” chapter,
if you need help with powering on the system.