Measure Proteins and Labels
Thermo Scientific
NanoDrop One User Guide
77
Calculations for Proteins and Labels Measurements
As with the other protein applications, Proteins & Labels
uses the
to correlate absorbance
with concentration based on the sample’s extinction
coefficient and pathlength.
This application offers six options (shown at right) for
selecting an appropriate extinction coefficient for each
measured sample, to be used in conjunction with Beer’s
Law to calculate sample concentration.
If the extinction coefficient of the sample is known,
choose the
+ MW (molar) or
1% (mass) option and
enter the value. Otherwise, calculate the extinction
coefficient or choose the option that best matches the
sample solution.
Tip
: Ideally, the extinction coefficient should be
determined empirically using a solution of the study
protein at a known concentration using the same buffer.
Available Options for Extinction Coefficient
•
1 Abs = 1 mg/mL
, where sample type and/or ext. coefficient is
unknown (produces rough estimate of protein concentration)
•
BSA
(Bovine Serum Albumin, 6.67 L/gm-cm)
•
IgG
(any mammalian antibody, 13.7 L/gm-cm)
•
Lysozyme
(egg white lysozyme, 26.4 L/gm-cm)
•
Other protein
(
+ MW), user-specified molar ext. coefficient
•
Other protein
(
1%), user-specified mass ext. coefficient
Note
: See
for details.
Calculated protein concentrations are based on the
absorbance value at 280 nm, the selected (or entered)
extinction coefficient and the sample pathlength. A
single-point baseline correction (or analysis correction)
may be applied.
Concentration is reported in mass units. Calculators are
available on the Internet to convert concentration from
mass to molar units based on sample sequence.
Measured Values
A280 absorbance
Note
: The absorbance value at 750 nm is subtracted from all wavelengths
in the spectrum. As a result, the absorbance at 750 nm is zero in the
displayed spectra. Also, for micro-volume absorbance measurements and
measurements taken with nonstandard (other than 10 mm) cuvettes, the
spectra are normalized to a 10 mm pathlength equivalent.
• Protein absorbance values are measured at 280 nm using the
750 nm-corrected and normalized spectrum. If Analysis Correction
and Dye Correction are not selected, this is the reported A280 value
and the value used to calculate protein concentration.
is selected, the 750-corrected, normalized and
analysis-corrected absorbance value at 280 nm is reported and used to
calculate protein concentration.
• If a Dye is used, the 750-corrected, normalized, analysis-corrected and
absorbance value at 280 nm is reported and used to
calculate protein concentration.
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 144: ......
Page 152: ......
Page 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Page 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Page 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...