305
When the System Arrives
When the instrument arrives, check the exterior of the shipping box for signs of damage. If
damage is apparent, contact us or your local distributor for instructions.
• Move the shipping box to the installation location at least 24 hours before installation.
• Keep the instrument upright at all times.
The warranty will not cover:
• Damage due to improper moving techniques.
• Damage due to removing the sealed plastic bag before the instrument has come to room
temperature.
•
WARNING
Avoid personal injury. If this equipment is used in a manner not specified in
the accompanying documentation, the protection provided by the equipment may be
impaired.
CAUTION
Avoid personal injury. Perform
only
those procedures described in the
documentation. If there are other problems, contact us. Any other service must be
performed by trained personnel.
CAUTION
Avoid shock hazard. Do not remove the cover of the instrument. All service to
the instrument must be performed by trained personnel.
NOTICE
• Inside the shipping box, the instrument is sealed in a plastic bag to keep the unit
dry.
• Allow 24 hours for the instrument to reach room temperature before opening the
bag. If the bag is opened before the instrument reaches room temperature,
moisture could condense on the optical components and cause permanent
damage.
Note
It is important to have all system utilities installed before the instrument arrives.
Utility installations must comply with all local building and safety codes.
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 144: ......
Page 152: ......
Page 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Page 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Page 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...