Measure Protein Lowry
112
NanoDrop One User Guide
Thermo Scientific
Protein assay kits and protocols
Follow the assay kit manufacturer’s recommendations for all standards and samples
(unknowns). Ensure each is subjected to the same timing and temperature throughout the
assay.
To measure Protein Lowry standards and samples
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument
arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new
laboratory wipe. For more information, see
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into
the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently
damage the quartz fiber optic cables.
To measure Protein Lowry standards and samples
1. From the Home screen, select the
Proteins
tab and tap
Protein Lowry
.
2. Specify a
and enter the
concentration of each standard
Tip
: For this assay, we recommend setting
Curve Type
to “2nd Order Polynomial.”
3. Measure blank:
–
pipette 2 μL DI H
2
O onto lower pedestal and lower arm, or insert DI H
2
O
blanking cuvette into cuvette holder
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 144: ......
Page 152: ......
Page 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Page 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Page 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...