Measure Proteins and Labels
76
NanoDrop One User Guide
Thermo Scientific
Related Topics
•
Detection Limits for Proteins and Labels Measurements
Detection limits and reproducibility specifications for purified BSA proteins and dyes that are
pre-defined in the software are provided
. The BSA lower detection limit and
reproducibility values apply to any protein sample type. The upper detection limits are
dependent on the
of the instrument and the sample’s extinction
coefficient.
To calculate upper detection limits for other (non-BSA) protein
sample types
To calculate upper detection limits in ng/μL for proteins, use the following equation:
(upper absorbance limit
instrumen
t
/mass extinction coefficient
sample
) * 10
For example, if the sample’s mass extinction coefficient at 280 nm is 6.67 for a 1%
(10 mg/mL) solution, the equation looks like this:
(550 / 6.67) * 10 = 824.6 (or ~825)