19
Maintenance
Instrument Diagnostics
306
NanoDrop One User Guide
Thermo Scientific
After the tenth measurement, a message indicates whether the instrument passed or failed
performance verification:
8. If the instrument failed, immediately repeat step 6 using ten 2 μL aliquots of the PV-1
solution.
9. When finished, tap
End Experiment
and
clean the pedestals with 3–5 μL DI H
O
.
After the test is completed, the results are available from the Data Viewer (see example
below). See
Manage identifiers on the instrument
for details.
To interpret performance verification results
If your instrument failed performance verification and you repeated ten measurements using
2 uL aliquots,