Maintenance
Instrument Diagnostics
Thermo Scientific
NanoDrop One User Guide
307
Pedestal Image Check
Run the Pedestal Image Check periodically to verify the instrument’s column sensor which
monitors for possible errors such as an empty column or bubbles in a sample. The Pedestal
Image Check can be used for routine quality control purposes. It also provides important
diagnostic information if a detection system component fails.
Supplies needed
• lint-free laboratory wipes
To run pedestal image check
1. Lift the instrument arm and clean the upper and lower pedestal with a new laboratory
wipe.
2. Lower the arm.
3. From the instrument home screen, tap
(Diagnostics) and then tap
Pedestal Image Check
.
4. Tap
Measure
.
The instrument runs a series of tests to check pedestal position and image quality. After
the measurements are completed, the results are displayed. A green check mark indicates
the instrument passed the Pedestal Image Check.
5. When finished, tap
End Experiment
.
To interpret pedestal image check results
If the Pedestal Image Check displays a yellow triangle instead of the green check mark, follow
the on-screen instructions to fix any possible problems. Then rerun the Pedestal Image Check.
If the instrument fails again,
Related Topics
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 144: ......
Page 152: ......
Page 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Page 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Page 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...