19
Maintenance
Maintaining the Pedestals
292
NanoDrop One User Guide
Thermo Scientific
Maintaining the Pedestals
The pedestals require periodic maintenance to maintain measurement integrity. Time lines
and procedures for cleaning and reconditioning the pedestals are provided below.
Cleaning the Pedestals
To avoid carryover and cross contamination, clean the pedestals before the first blank or
sample measurement and at the end of each measurement. Additional cleaning (see below) or
may be required for periodic maintenance.
NOTICE
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently
damage the quartz fiber optic cables.
• To prevent damage from spills, keep containers of liquids away from the instrument.
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into
the instrument and may cause permanent damage.
• Do not attempt to remove the diaphragm around the lower pedestal as it is
permanently affixed to the instrument.
• Do not allow HCl, alcohol, bleach, acetone or any other solvent to remain on the
diaphragm for more than one minute or it may loosen the seals. If the diaphragm
becomes loose,
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 144: ......
Page 152: ......
Page 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Page 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Page 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...