13
Measure Protein Pierce 660
Thermo Scientific
NanoDrop One User Guide
127
Settings for Protein Pierce 660 Measurements
To show the Protein Pierce 660 settings, from the Protein Pierce 660 measurement screen, tap
>
Protein Pierce 660 Setup.
Related Topics
Note
You can edit the Curve Type setting when measuring standards by changing the list
box at the top of the application measurement screen. You can edit the concentration
value for a standard from the application setup screen. After the first sample measurement,
these settings cannot be changed.
Setting
Description
Curve Type
Specify type of equation used to create standard curve from standard concentration values.
Available options:
–
Linear
: Draws the linear least squares line through all measured standards (requires
reference measurement and at least one standard)
–
Interpolation
: Draws a series of straight lines to connect all measured standards
(requires reference measurement and at least one standard)
–
2
nd
order polynomial
: Draws the 2
nd
order least squares polynomial using all
measured standards (requires reference measurement and at least two standards)
–
3
rd
order polynomial
: Draws the 3
rd
order least squares polynomial using all
measured standards (requires reference measurement and at least three standards)
Replicates
Enter number of measurements of the reference or the same standard or sample that are
averaged together to produce its associated concentration value.
Note
: Replicates setting cannot be changed after the first standard has been measured.
Standards
Enter actual concentration value of each standard.
Note
: Concentration values can be entered in any order but the standards must be measured
in the order in which they were entered.
If you also want to enter previously measured absorbance values for the standards, select this
check box:
and then enter absorbance values for all the standards.
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 144: ......
Page 152: ......
Page 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Page 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Page 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...