Measure Kinetics
Measure Kinetics
The NanoDrop One
C
model instrument can be used to make time-based kinetic
measurements on samples in cuvettes. Up to 3 wavelengths between 190 nm and 850 nm can
be designated for continuous absorbance monitoring at user-defined intervals in up to
5 stages. Cuvette measurements offer an extended lower
and an optional 37 °C
heater and micro-stirrer.
To make kinetic measurements
Make time-based kinetic
measurements using the cuvette
holder (NanoDrop One
C
model
instruments only).
Note
The instrument arm can be up during cuvette measurements, which allows you to
add reagents to the sample solution if desired.
NOTICE
• To prevent damage from spills, keep containers of liquids away from the instrument.
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into
the instrument and may cause permanent damage.