Instrument Diagnostics
304
NanoDrop One User Guide
Thermo Scientific
4. Pipette 1 μL DI H
2
O onto the lower pedestal, lower the arm and tap
Blank
.
5. Lift the arm and clean both pedestals with a new wipe.
6. Pipette 1 μL PV-1 solution onto the lower pedestal and start the sample measurement:
is On, lower arm
• If Auto-Measure is off, lower arm and tap
Measure
After the measurement, the software displays the results. Here is an example of the
performance verification result screen.
7. Repeat step 6 to measure the PV-1 solution nine more times using a new 1 μL aliquot for
each measurement and cleaning both pedestals after each measurement.
Note
Vigorously shake the ampoule of PV-1 solution, allow the liquid to collect in
the bottom portion of the ampoule and then follow standard practices to open it.
Swipe screen left
to
view detailed results
Measured pathlengths
Drag tab down/up
to
see more/less data