Measure a Micro-Volume Sample
190
NanoDrop One User Guide
Thermo Scientific
Recommended sample volumes
Pipetting Samples
• Use
to
ensure proper liquid column
formation.
• Use calibrated precision pipettor (0–
2 μL volume range) with well-fitting,
low-retention precision tips to apply
sample material to instrument for
measurement.
If using low accuracy (0-10 μL)
pipettor, use 2 μL sample volumes.
• Use new tip for each blank and sample
aliquot.
• Use new aliquot of sample for each
measurement.
• If solvents are used, make sure they are
compatible with the pedestals. (see
“Compatible Solvents” in
Application
Sample Volume
Nucleic acid (aqueous solution)
1 μL
a
a
Use 2 μL for samples that contain materials that may reduce surface tension such as a surfactant.
Purified protein
2 μL
Other protein applications such as Bradford or BCA
2 μL
Microbial cell suspensions
2 μL