13
Measure Protein Pierce 660
Thermo Scientific
NanoDrop One User Guide
125
Protein Pierce 660 standard curve screen
The standard curve screen shows graphically the relationship between the measured standards,
the calculated standard curve, and the measured absorbance and calculated concentration for
a selected sample. A horizontal line connects the sample absorbance value on the Y-axis to the
standard curve. A vertical line connects that point to the sample concentration value on the
X-axis.
The R
2
value indicates how well the standard curve fits the standard data points (1.0 is a
perfect fit; that is, all points lie exactly on the curve).
Tap row
to
select sample
and update
graph;
tap more
rows
to overlay
abs/conc lines
for up to five
samples.
Press
and hold
sample row for
measurement
details.
Drag tab
down/up
to see
more/less
sample data
Pinch and zoom
to adjust axes;
double-tap to reset
Standard curve
Menu of options;
tap
to open
Protein concentration
for selected sample
Absorbance
at 660 nm
Page control;
swipe screen left or
right
to view next or previous screen
R
2
value (1.0
equals perfect fit)
Red squares indicate
sample data points
White circles indicate
standard data points
Selected sample
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 144: ......
Page 152: ......
Page 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Page 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Page 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...