4
Measure Microarray
Thermo Scientific
NanoDrop One User Guide
31
Add or remove a dye
To add or remove a dye from the Dye1 or Dye2 drop-down list in
or
–
select or deselect corresponding checkbox
Add custom dye
–
tap
to show New Dye box
–
enter unique
Name
for new dye (tap field to display keyboard, tap
Done
key to close
keyboard)
–
select default
Unit
that will be used to display dye concentration
–
enter dye’s
Extinction Coefficient
(or molar absorptivity constant) in L/mole-cm
(typically provided by dye manufacturer)
–
specify
Wavelength
in nm (between 450 nm and 700 nm) that will be used to
measure dye’s absorbance
–
specify dye’s correction values at 260 nm and 280 nm
–
tap
Add Dye
When a custom dye is selected before a measurement, the dye’s absorbance and
concentration values are reported and the corrections are applied to the measured sample
absorbance values, and to the resulting sample concentrations and purity ratios.
Edit custom dye
–
tap to select custom dye
–
tap
Note
To determine dye correction values (if not available from dye manufacturer):
–
use instrument to measure pure dye and note absorbance at 260 nm, 280 nm
and at analysis wavelength for dye (see above)
–
calculate ratio of A
260
/A
dye wavelength
and enter that value for 260 nm
Correction
–
calculate ratio of A
280
/A
dye wavelength
and enter that value for 280 nm
Correction
Tip
Dyes pre-defined in the software cannot be edited.
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 144: ......
Page 152: ......
Page 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Page 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Page 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...