Dionex Reagent-Free Controller
72
Doc. 031880-04 1/12
DX-120 Pump Pressure Alarm Setup
Using the DIP switch labeled switch 1 on Figure 62, set the high pressure
alarm to 24.1 MPa (3500 psi) (position 5 = ON) and set the low pressure
alarm to 2.0 MPa (250 psi) (position 8 = ON).
Figure 69. DX-120 High and Low Pressure
Alarm DIP Switch Settings
If you are installing a Dionex RFC-10, go on to Section B.6. If you are
installing a Dionex RFC-30, go on to Section B.7.
B.6
Dionex RFC-10 Initial Setup
NOTE See Section B.7 for Dionex RFC-30 setup.
Before beginning operation of the Dionex RFC-10, refer to the suppressor
manual for the recommended current setting for your application.
DIP Switch 1
24.1 Mpa (3500 psi)(Pos. 5=ON)
2.0 Mpa (250 psi)(Pos. 8=ON)
High Pressure Alarm
Low Pressure Alarm
Steps for initial setup of the Dionex
RFC-10
Note
: The maximum current that can be
applied to the suppressor depends on the
suppressor type and the flow rate.
2. Set pum p flow
3. Set suppressor current
1. Set suppressor type