Dionex Reagent-Free Controller
60
Doc. 031880-04 1/12
9. Turn the new holder
upside down (see
Figure 56) and connect
the liquid lines to the
CR-TC.
NOTE To help retain the
fittings and ferrules on
the lines, remove the
union from each line
just before connecting
the line to a port.
Figure 56. EGC Holder (upside down)
10. Check that none of the lines are caught under the CR-TC and then
push the CR-TC onto the metal stud inside the EGC holder (see
Figure 56). Push down until the CR-TC snaps in place.
11. Feed the CR-TC cable through the slot on the side of the EGC holder
12. Turn the EGC holder right side up.
13. Reconnect the
EGC IN
and
EGC OUT
lines to the EGC inlet and outlet
ports.
14. Turn over the EGC (fittings facing down).
Shake the EGC
vigorously, and tap it with the palm of your hand 10 to 15 times.
Check that all trapped bubbles are dislodged.
15. Reinstall the EGC in the holder.
16. Remove the cap on the luer adapter by twisting.
Save the cap
.
17. Locate the EGC vent line tubing, label, tubing clips, and barbed luer
lock, supplied with the holder. Slide the label onto the tubing and
push the tubing onto the luer lock.
18. Twist the luer lock and EGC vent line into the luer adapter
Unions
Metal stud for mounting the CR-TC