Service Procedures
Doc. 031880-04 1/12
61
19. Remove the paper backing from the black tubing clips and press the
clips onto the sides of the EGC holder (see Figure 57). Attach the
vent line to the clips.
Figure 57. EGC Vent Line and Tubing Clips
20. Route the vent line to the waste container and insert it down into the
gas separator waste tube about 4 to 6 inches.
21. Reconnect the CR-TC and EGC cables.
22. Connect the holder liquid lines to the pump, suppressor, injection
valve, and gas separator waste tube assembly.
23. Turn on the power to the Dionex RFC.
Tubing
Clips