Service Procedures
Doc. 031880-04 1/12
75
Set the Initial Flow Rate
1. Press
three times to display the pump flow screen.
Figure 79. Pump Flow Screen
2. Press
ENTER
.
3. Press
to select a flow rate of
1.00 mL/min.
Note:
This is the flow rate required for initial set up of
the EGC. You will set the application flow rate later.
Figure 80. 1.0 mL Flow Rate Selected
4. Press
ENTER
.
Set the Initial Eluent Concentration
1. Press
to display the EGC concentration screen.
Figure 81. EGC Concentration Screen
2. Press
ENTER
.
3. Press
to select an eluent concentration of 50 mM.
Note:
This is the
concentration required for initial set up of the EGC. You will set the
application concentration later.
Figure 82. 50 mM Concentration Selected
4. Press
ENTER
.
5. Go on to Section B.8.