Dionex Reagent-Free Controller
44
Doc. 031880-04 1/12
•
ERR: NO BOARD PRESENT [150]
The Dionex RFC does not recognize the installed electronics board(s)
or no board is installed.
To troubleshoot:
•
If this error appears, call Thermo Fisher Scientific for assistance.
The Dionex RFC electronics components cannot be serviced by
the user.
•
ERR: BOARD IS BAD—SUPPRESSOR [152]
The suppressor controller electronics board is malfunctioning.
To troubleshoot:
•
Call Thermo Fisher Scientific for assistance. The Dionex RFC
electronics components cannot be serviced by the user.
•
ERR: BOARD IS BAD—EGC [153]
The EGC controller electronics board is malfunctioning.
To troubleshoot:
•
Call Thermo Fisher Scientific for assistance. The Dionex RFC
electronics components cannot be serviced by the user.
•
WRN: NEW BOARD FOUND
MUST CALIBRATE [160]
This warning occurs when a new electronics board is installed.
To troubleshoot:
•
If this warning occurs when a new board has not been installed,
first try calibrating the Dionex RFC (see Section 6.2).
•
If the warning persists, call Thermo Fisher Scientific for
assistance. The Dionex RFC electronics components cannot be
serviced by the user.