Dionex Reagent-Free Controller
62
Doc. 031880-04 1/12
A.
Dionex RFC Specifications
Input Voltage Range
90 – 264 Vac, 47 – 63 Hz
Maximum AC Power Draw
45 W
Maximum AC Current Draw
0.75 A
Typical AC Current Draw
Dionex RFC-10:
0.4 A
Dionex RFC-30:
0.5 A
Output Operating Range
AES:
CC
a
:
0 – 150 mA, 0 – 50 VDC
SRS:
CC:
0 – 500 mA, 0 – 10 VDC
EG-1/EG-2:
CC:
0 – 320 mA, 0 – 45 VDC
CR-TC:
CV
b
: 24 VDC, 0 – 170 mA
a. CC=constant current
b. CV=constant voltage
Ripple and Noise
AES/SRS/EG
: CC: less than 0.01 mA rms and
0.020 mA peak-to-peak
CR-TC
: CV: less than 200
V rms and 2 mV
peak-to-peak
Output Drift
AES/SRS/EG
: CC: less than 0.1% total drift for 8
hours after an initial warm-up of 10 minutes
CR-TC
: CV: less than 0.1% total drift for 8 hours
after an initial warm-up of 10 minutes
Overload Protection
A continuously acting constant current circuit
protects the power supply for all overloads,
including a direct short on its power output
TTL Connectors
TTL inputs operate from 0 to 5 VDC level. Inputs
are active low (less than or equal to 0.5 VDC)
Fuses
Two IEC 127 fast-blow fuses rated 3.15 amp
Temperature Range
0 – 40 °C (32 – 104 °F); fan cooling is employed
Weight (net)
2.9 kg (6.5 lbs)