Service Procedures
Doc. 031880-04 1/12
81
2. Disconnect the coiled black
EGC IN
line from the union that connects
it to the
EGC OUT
line. Connect
EGC IN
to the EGC
INLET
Figure 91. EGC Inlet Connection
3. Locate the yellow 0.5 mL/min, 1000 psi (7 MPa) pressure restrictor
coil (P/N 053765) in the Dionex RFC-30 Ship Kit. Connect this coil
to the EGC
OUTLET
Note:
This is a temporary
connection, which is used only during the EGC conditioning
procedure described on page 83.
Figure 92. EGC Outlet Connection for Conditioning the EGC
INLET Port
EGC IN
Pressure Restrictor
Coil
OUTLET
Port