C-Link Protocol Commands
Calibration
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
B-25
Send:
set sp field 1 34
Receive:
set sp field 1 34 ok
stream per
This command reports the currently set time interval in seconds for
streaming data.
Send:
stream
per
Receive: stream
per
10
set stream per
number value
number value
= | 1 | 2 | 5 | 10 | 20 | 30 | 60 | 90 |120 | 180 | 240 | 300 |
This command sets the time interval between two consecutive streaming
data strings to
number value
in seconds. The example below sets the
number value to 10 seconds.
Send:
set stream per 10
Receive:
set stream per 10 ok
stream time
This command reports if the streaming data string will have a time stamp
attached to it or not, according to
Table B–5
.
Send:
stream
time
Receive:
stream time 0
set stream time
value
This command enables
value
, where
value
is to attach or disable time stamp
to streaming data string, according to
Table B–5
. The example below
attaches a time stamp to streaming data.
Send:
set stream time 0
Receive:
set stream time 0 ok
Table B–5.
Stream Time Values
Value
Stream Time
00
Attaches time stamp to streaming data string
01
Disables time stamp to streaming data string
set cal no bkg
set cal nox bkg
set cal pre bkg
These commands will auto-calibrate the NO, NO
x
, and prereactor
backgrounds. If the instrument is set to manual NO
x
mode, the response to
Calibration