Operation
Instrument Controls Menu
3-44
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > Select Relay >
Instrument State
.
CHOOSE SIGNAL TYPE:
>ALARMS
NON-ALARM
RANGE AVG DIAGS ALARM
The Alarms status screen allows the user to select the alarm status for the
selected relay output. The selected item is shown by “<--” after it. The I/O
board status alarm is only present if the I/O expansion board is installed.
The zero and span check/calibration alarms are only present if the
zero/span valve option is installed and the instrument is operating in
manual mode.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > Select Relay > Instrument State >
Alarms
.
ALARM STATUS ITEMS:
>NONE
GEN ALARM <--
NO CONC MAX
NO CONC MIN
NO2 CONC MAX
NO2 CONC MIN
NOx CONC MAX
ª
RANGE AVG DIAGS ALARM
NOx CONC MIN
INT TEMP
CHAMB TEMP
COOLER TEMP
NO2 CV TMEP
PRESSURE
FLOW
OZONE FLOW
MB STATUS
MIB STATUS
I/O BD STATUS
CONC ALARM
ZERO CHK/CAL
SPAN CHK/CAL
Alarms