Servicing
Firmware Updates
7-4
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
Figure 7–1.
Properly Grounded Antistatic Wrist Strap
The firmware can be updated by the user in the field via the serial port or
over the Ethernet. This includes both the main processor firmware and the
firmware in all low-level processors. Refer to the
iPort
manual for the
firmware update procedure.
If the Service Menu is not displayed on the Main Menu, use the following
procedure to display it.
1.
At the Main Menu, press
to scroll to Instrument Controls >
press
>
to scroll to
Service Mode
> and press
.
The Service Mode screen appears.
2.
Press
to toggle the Service Mode to ON.
3.
Press
>
to return to the Main Menu.
4.
Return to the procedure.
Table 7–1
lists the replacement parts for the Model 42
i
Trace Level major
subassemblies. Refer to
Figure 7–2
to identify the component location.
Table 7–1.
Model 42
i
Trace Level Replacement Parts
Part Number
Description
100480-00
Front Panel Pushbutton Board
101491-04 Processor
Board
100533-00 Motherboard
100539-00
Digital Output Board
100542-00
I/O Expansion Board (Optional)
102340-00
Front Panel Connector Board
Firmware
Updates
Accessing the
Service Mode
Replacement
Parts List