Servicing
Ozonator Transformer Replacement
7-26
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
Figure 7–13.
Replacing the Ozonator Assembly
5.
Remove the two screws securing the ozonator to the ozonator bracket.
6.
Unplug the ozonator from the ozonator transformer by lifting the
ozonator straight up.
7.
To install the ozonator, follow the previous steps in reverse.
8.
Re-install the measurement bench. Refer to “
Removing the
Measurement Bench and Lowering the Partition Panel
” in this chapter.
Use the following procedure to replace the ozonator transformer (
Figure 7–
13
).
Equipment Required:
Ozonator transformer
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn instrument OFF, unplug the power cord, and remove the cover.
Ozonator Transformer
Replacement
Ozonator – Side View
Ozonator Bracket Screws (2)
Ozonator Transformer
Ozonator
Ozonator Transformer
Screws (4)