Operation
Service Menu
3-78
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
CALIBRATE FLOW ZERO:
CURRENTLY: 0.004 LPM
SET TO: 0.000 LPM ?
DISCONNECT PUMP AND
SAVE CURRENT FLOW
RANGE AVG DIAGS ALARM
The Calibrate Flow Span screen allows the user to view and set the flow
sensor calibration span point.
Note
An independent flow sensor is required to read the flow, then the
operator enters the flow value on this screen to perform the calibration.
▲
●
In the Main Menu, choose Service > Flow Calibration >
Span
.
CALIBRATE FLOW SPAN:
CURRENTLY: 1.000 LPM
SET TO: 0.80
0
LPM ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Restore Default Flow Calibration screen allows the user to reset the
flow calibration configuration values to factory defaults.
●
In the Main Menu, choose Service > Select Flow Calibration A or B >
Set Defaults
.
Calibrate Flow Span
Restore Default Flow
Calibration
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM