C-Link Protocol Commands
Communications Configuration
B-38
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
set time
hh:mm:ss
hh
= hours
mm
= minutes
ss
= seconds
This command sets the internal clock (24-hour time). The example below
sets the internal time to 2:15 pm.
Note
If seconds are omitted, the seconds default to 00.
▲
Send:
set
time
14:15
Receive:
set time 14:15 ok
addr dns
This command reports the TCP/IP address for the domain name server.
Send:
addr
dns
Receive:
addr dns 192.168.1.1
set addr dns
address
This command sets the dns
address
, where
address
consists of four numbers
ranging from 0-255 inclusive, separated by “.”.
Send:
set addr dns 192.168.1.1
Receive:
set addr dns 192.168.1.1 ok
addr gw
This command reports the default TCP/IP gateway address.
Send:
addr
gw
Receive:
addr gw 192.168.1.1
set addr gw
address
This command sets the default gateway
address
, where
address
consists of
four numbers ranging from 0-255 inclusive, separated by “.”.
Send:
set addr gw 192.168.1.1
Receive:
set addr gw 192.168.1.1 ok
addr ip
This command reports the IP address of the analyzer.
Send:
addr
ip
Receive:
addr ip 192.168.1.15
set addr ip
address
This command sets the analyzer’s IP
address
, where
address
consists of four
numbers ranging from 0-255 inclusive, separated by “.”.
Communications
Configuration