Installation
Setup Procedure
2-4
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
2.
Connect the air dryer to the DRY AIR bulkhead.
3.
Connect the
pump vacuum port (inlet) to the EXHAUST bulkhead
(figure 2-4).
Figure 2–4.
Twin-Head Vacuum Pump Installation
4.
Connect the pump exhaust to a suitable vent or charcoal scrubber.
Equipment Damage
DO NOT connect the pump exhaust port to the
EXHAUST bulkhead. This can cause permanent damage to both the
cooler and the photomultiplier tube.
▲
5.
Connect a suitable recording device to the rear panel connector. For
detailed information about connecting to the instrument, refer to:
“
Connecting External Devices
” on page
2-5
“
External Device Connection Components
” on page
7-6
“
Terminal Block and Cable Kits
” on page
9-2
“
Instrument Controls Menu
” on page
3-25
For detailed information about troubleshooting a connection, refer to
“
Analog Output Testing
” on page
7-30
.