Operation
Alarms Menu
3-66
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
The Minimum Cooler Temperature alarm limit screen is used to change
the minimum cooler temperature alarm limit. The minimum and
maximum cooler temperature screens function the same way.
●
In the Main Menu, choose Alarms > Cooler Temp >
Min
or
Max
.
COOLER TEMPERATURE:
ACTUAL MIN -20.0
o
C
SET MIN TO: -10.0
o
C ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Converter Temperature screen displays the current converter
temperature and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 300 to 700 °C. The actual alarm setpoints should
be set for the installed converter. If the converter temperature reading goes
beyond either the minimum or maximum limit, an alarm is activated. The
word “ALARM” appears in the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Converter Temp
.
CONVERTER TEMPERATURE:
ACTUAL 320.7
o
C
>MIN 300.0
o
C
MAX 350.0
o
C
RANGE AVG DIAGS ALARM
The Minimum Converter Temperature alarm limit screen is used to change
the minimum converter temperature alarm limit. The minimum and
maximum converter temperature screens function the same way.
●
In the Main Menu, choose Alarms > Converter Temp >
Min
or
Max
.
Min and Max Cooler
Temperature Limits
Converter Temperature
Min and Max Converter
Temperature Limits