Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-27
PMT SUPPLY:
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
The Auto/Manual Mode screen allows selection of the automatic mode
(NO/NO
x
), NO mode (manual NO), NO
x
mode (manual NO
x
) or
prereactor mode (manual prereact). The auto cycle mode switches the
mode solenoid valve automatically on a 10 second cycle so that NO, NO
2
,
and NO
x
concentrations are determined. The manual NO mode puts the
mode solenoid valve into the open position so that the sample gas bypasses
the NO
2
-to-NO converter. Therefore, only the NO concentration is
determined. The manual NOx mode puts the mode solenoid valve into the
closed position so that the sample gas passes through the NO
2
-to-NO
converter. Therefore, only the NO
x
concentration is determined. In the
manual modes, additional averaging times of 1, 2, and 5 seconds are
available from the Averaging Times screen.
●
In the Main Menu, choose Instrument Controls >
Auto/Manual
Mode
.
MODE:
CURRENTLY: MANUAL NO
SET TO: NO/NOX ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Datalogging Settings menu deals with datalogging.
●
In the Main Menu, choose Instrument Controls >
Datalogging
Settings
.
Auto/Manual Mode
Datalogging Settings