Operation
Alarms Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-67
CONVERTER TEMPERATURE:
ACTUAL MIN 300.0
o
C
SET MIN TO: 310.0
o
C ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Pressure screen displays the current reaction chamber pressure reading
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 200 to 450 mmHg. If the pressure reading goes beyond either
the minimum or maximum limit, an alarm is activated. The word
“ALARM” appears in the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Pressure
.
PRESSURE:
ACTUAL 232.5 mmHg
>MIN 200.0 mmHg
MAX 450.0 mmHg
RANGE AVG DIAGS ALARM
The Minimum Pressure alarm limit screen is used to change the minimum
temperature alarm limit. The minimum and maximum pressure screens
function the same way.
●
In the Main Menu, choose Alarms > Pressure >
Min
or
Max
.
PRESSURE:
ACTUAL MIN: 200.0 mmHg
SET MIN TO: 210.0 mmHg?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Pressure
Min and Max Pressure
Limits