Installation
Startup
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
2-9
Use the following procedure when starting the instrument.
1.
Turn the power ON.
2.
Allow 90 minutes for the instrument to stabilize.
3.
Set instrument parameters such as operating ranges and averaging times
to appropriate settings. For more information about instrument
parameters, see the “
Operation
” chapter.
4.
Before beginning the actual monitoring, perform a multipoint
calibration as described in the “
Calibration
” chapter.
Startup