Servicing
Ozonator Assembly Replacement
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
7-25
Use the following procedure to replace the ozonator assembly (
Figure 7–
13
).
Equipment Required:
Ozonator assembly
Wrench, 5/8-inch
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Refer to “
Removing the Measurement Bench and Lowering the
Partition Panel
” in this chapter to lower the partition panel, then
proceed to the next step below.
2.
Carefully disconnect the plumbing at the glass inlet and outlet of the
ozonator.
3.
Disconnect the stainless steel tubing from the flow transducer.
4.
Loosen the four captive screws securing the ozonator bracket to the
floor plate.
Ozonator Assembly
Replacement
Captive Screws (4)
Flow Transducer
Ozonator – Top View