Servicing
Fuse Replacement
7-40
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
Note
If Service Mode is not displayed, refer to “
Accessing the Service
Mode
” on page
7-4
, then return to the beginning of this step.
▲
4.
Wait at least 10 seconds for the ambient reading to stabilize, use
and
to enter the known temperature, and
press
to save the temperature value.
5.
Install the cover.
Use the following procedure to replace the fuse.
Equipment Required:
Replacement fuses:
250VAC, 4 Amp, SlowBlow (for 100VAC and 110VAC models)
250VAC, 2 Amp, SlowBlow (for 220-240VAC models)
1.
Turn instrument OFF and unplug the power cord.
2.
Remove fuse drawer, located on the AC power connector.
3.
If either fuse is blown, replace both fuses.
4.
Insert fuse drawer and reconnect power cord.
Use the following procedure to replace the ammonia scrubber (
Figure 7–
18
).
Equipment Required:
Ammonia scrubber
Nut driver, 3/16-inch
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn instrument OFF, unplug the power cord, and remove the cover.
2.
Unscrew the Teflon tubing at both ends of the scrubber.
Fuse Replacement
Scrubber
Replacement