Operation
Calibration Factors Menu
3-16
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
the dual and auto range modes have two averaging times (high and low).
The Averaging Time screen functions the same way in the single, dual, and
auto range modes. The following averaging times are available: 10, 20, 30,
60, 90, 120, 180, 240, and 300 seconds. Additional averaging times are
available in NO and NO
x
mode, see “
Auto/Manual Mode
” later in this
chapter.
●
In the Main Menu, choose
Averaging Time
.
AVERAGING TIME:
CURRENTLY: 30 SEC
SET TO: 10 SEC ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Calibration factors are used to correct the NO, NO
2
, and NO
x
concentrations readings that the instrument generates using its own
internal calibration data. The Calibration Factors menu displays the
calibration factors. The screens below show the calibration factors menu in
single mode and dual/auto range modes. The only difference between the
screens are the words “HI” and “LO” to indicate which range is displayed.
Normally, the instrument is calibrated automatically using the Calibration
menu described in “
Calibration Menu
” later in this chapter. However, the
instrument can also be calibrated manually using the Calibration Factors
menu.
To manually calibrate the instrument, see “
NO, NOx, and Prereactor
Backgrounds
”, “
NO, NO2, and NOx Coefficients
” and “Calibration
Pressure” below for more information.
●
In the Main Menu, choose
Calibration Factors
.
Calibration Factors
Menu