C-Link Protocol Commands
Hardware Configuration
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
B-37
set ozonator safety
onoff
These commands set the ozonator safety
on
or
off
. The example below sets
the ozonator safety off.
Send:
set ozonator safety off
Receive:
set ozonator safety off ok
ozonator status
This command reports the status of the ozonator and safety. The example
below reports that the ozonator is off.
Send:
ozonator
status
Receive: ozonator
status
off
pmt status
This command reports the status of the PMT on or off. The example
below reports that the PMT is on.
Send:
pmt
status
Receive: pmt
status
on
set pmt
onoff
These commands set the PMT
on
or
off
. The example below turns the
PMT off.
Send:
set
pmt
off
Receive:
set pmt off ok
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example below saves the parameters to FLASH
memory.
Send:
set
save
params
Receive:
set save params ok
time
This command reports the current time (24-hour time). The example
below reports that the internal time is 2:15:30 pm.
Send:
time
Receive: time
14:15:30