C-Link Protocol Commands
Diagnostics
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
B-17
These commands set the NO, NO
2
, NO
x,
and prereactor concentration
alarm minimum
value
, where
value
is set to either floor or ceiling,
according to
Table B–3
. The example below sets the NO concentration
minimum alarm trigger to ceiling.
Send:
set alarm trig conc no 1
Receive:
set alarm trig conc no 1 ok
Table B–3.
Alarm Trigger Values
Value
Alarm Trigger
00 Floor
01 Ceiling
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag
volt
mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3, positive 18 IR, positive 18
MOT, and VBIAS. Each voltage value is separated by a space.
Send:
diag
volt
mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2 17.9 17.9
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag
volt
iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2
clr records
This command will clear all long and short records that have been saved.
Send:
clear
records
Receive:
clear records ok
Diagnostics
Datalogging